• Critical care medicine · Jan 2012

    Randomized Controlled Trial

    A clinical assessment of the Mucus Shaver: a device to keep the endotracheal tube free from secretions.

    • Lorenzo Berra, Andrea Coppadoro, Edward A Bittner, Theodor Kolobow, Patrice Laquerriere, Joshua R Pohlmann, Simone Bramati, Joel Moss, and Antonio Pesenti.
    • Department of Anesthesia, Critical Care, and Pain Medicine, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts, USA. lberra@partners.org
    • Crit. Care Med. 2012 Jan 1; 40 (1): 119-24.

    Objective: We evaluated a new device designed to clean the endotracheal tube in mechanically ventilated patients, the Mucus Shaver.Design: Prospective, randomized trial.Setting: University hospital intensive care unit.Patients: We enrolled 24 patients expected to remain ventilated for >72 hrs.Interventions: The Mucus Shaver is a concentric inflatable catheter for the removal of mucus and secretions from the interior surface of the endotracheal tube. The Mucus Shaver is advanced to the distal endotracheal tube tip, inflated, and subsequently withdrawn over a period of 3-5 secs. Patients were prospectively randomized within 2 hrs of intubation to receive standard endotracheal tube suctioning treatment or standard suctioning plus Mucus Shaver use until extubation.Measurements And Main Results: During the study period, demographic data, recent medical history, adverse events, and staff evaluation of the Mucus Shaver were recorded. At extubation, each endotracheal tube was removed, cultured, and analyzed by scanning electron microscopy. Twelve patients were assigned to the study group and 12 were assigned to the control group. No adverse events related to the use of the Mucus Shaver were observed. At extubation, only one endotracheal tube from the Mucus Shaver group was colonized, whereas in the control group ten endotracheal tubes were colonized (8% vs. 83%; p < .001). Scanning electron microscopy showed little secretions on the endotracheal tubes from the study group, whereas thick bacterial deposits were present on all the endotracheal tubes from the control group (p < .001 by Fisher exact test, using a maximum biofilm thickness of 30 μm as cut-off). The nursing staff was satisfied by the overall safety, feasibility, and efficacy of the Mucus Shaver.Conclusions: The Mucus Shaver is a safe, feasible, and efficient device for endotracheal tube cleaning in the clinical setting. The Mucus Shaver is helpful in preventing endotracheal tube colonization by potentially harmful microorganisms.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.