JACC Cardiovasc Interv · Jan 2014
Multicenter Study Clinical TrialA next-generation bioresorbable coronary scaffold system: from bench to first clinical evaluation: 6- and 12-month clinical and multimodality imaging results.
- Stefan Verheye, John A Ormiston, James Stewart, Mark Webster, Elias Sanidas, Ricardo Costa, J Ribamar Costa, Daniel Chamie, Andrea S Abizaid, Ibraim Pinto, Lynn Morrison, Sara Toyloy, Vinayak Bhat, John Yan, and Alexandre Abizaid.
- Antwerp Cardiovascular Center, ZNA Middelheim, Antwerp, Belgium. Electronic address: stefan.verheye@gmail.com.
- JACC Cardiovasc Interv. 2014 Jan 1;7(1):89-99.
ObjectivesThis study sought to perform clinical and imaging assessments of the DESolve Bioresorbable Coronary Scaffold (BCS).BackgroundBCS, which is drug eluting, may have potential advantages compared with conventional metallic drug-eluting stents. The DESolve system, designed to provide vessel support and neointimal suppression, combines a poly-l-lactic acid-based scaffold with the antiproliferative myolimus.MethodsThe DESolve First-in-Man (a non-randomized, consecutive enrollment evaluation of the DESolve myolimus eluting bioresorbable coronary stent in the treatment of patients with de novo native coronary artery lesions) trial was a prospective multicenter study enrolling 16 patients eligible for treatment. The principal safety endpoint was a composite of cardiac death, myocardial infarction, and clinically indicated target lesion revascularization. The principal imaging endpoint was in-scaffold late lumen loss (LLL) assessed by quantitative coronary angiography (QCA) at 6 months. Intravascular ultrasound (IVUS) and optical coherence tomography (OCT) imaging was performed at baseline and 6 months; multislice computed tomography (MSCT) was performed at 12 months.ResultsAcute procedural success was achieved in 15 of 15 patients receiving a study scaffold. At 12 months, there was no scaffold thrombosis and no major adverse cardiac events directly attributable to the scaffold. At 6 months, in-scaffold LLL (by QCA) was 0.19 ± 0.19 mm; neointimal volume (by IVUS) was 7.19 ± 3.56%, with no evidence of scaffold recoil or late malapposition. Findings were confirmed with OCT and showed uniform, thin neointimal coverage (0.12 ± 0.04 mm). At 12 months, MSCT demonstrated excellent vessel patency.ConclusionsThis study demonstrated the feasibility and efficacy of the DESolve BCS. Results showing low in-scaffold LLL, low % neointimal volume at 6 months, no chronic recoil, and maintenance of lumen patency at 12 months prompt further study. (DESolve First-in-Man; EudraCT number 2011-000027-32).Copyright © 2014 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.