• Europace · Jan 2015

    Randomized Controlled Trial Multicenter Study

    The NO Regular Defibrillation testing In Cardioverter Defibrillator Implantation (NORDIC ICD) trial: concept and design of a randomized, controlled trial of intra-operative defibrillation testing during de novo defibrillator implantation.

    • Dietmar Bänsch, Hendrik Bonnemeier, Johan Brandt, Frank Bode, Jesper Hastrup Svendsen, Angelika Felk, Tino Hauser, Karl Wegscheider, and NORDIC ICD Trial Investigators.
    • Heart Center Rostock, Department of Internal Medicine I, Divisions of Cardiology, University Hospital Rostock, Ernst-Heydemann-Str. 6, 18057 Rostock, Germany dietmar.baensch@med.uni-rostock.de.
    • Europace. 2015 Jan 1;17(1):142-7.

    AimsAlthough defibrillation (DF) testing is still considered a standard procedure during implantable cardioverter-defibrillator (ICD) insertion and has been an essential element of all trials that demonstrated the survival benefit of ICD therapy, there are no large randomized clinical trials demonstrating that DF testing improves clinical outcome and if the outcome would remain the same by omitting DF testing.Methods And ResultsBetween February 2011 and July 2013, we randomly assigned 1077 patients to ICD implantation with (n = 540) or without (n = 537) DF testing. The intra-operative DF testing was standardized across all participating centres. After inducing a fast ventricular tachycardia (VT) with a heart rate ≥240 b.p.m. or ventricular fibrillation (VF) with a low-energy T-wave shock, DF was attempted with an initial 15 J shock. If the shock reversed the VT or VF, DF testing was considered successful and terminated. If unsuccessful, two effective 24 J shocks were administered. If DF was unsuccessful, the system was reconfigured and another DF testing was performed. An ICD shock energy of 40 J had to be programmed in all patients for treatment of spontaneous VT/VF episodes. The primary endpoint was the average efficacy of the first ICD shock for all true VT/VF episodes in each patient during follow-up. The secondary endpoints included the frequency of system revisions, total fluoroscopy, implantation time, procedural serious adverse events, and all-cause, cardiac, and arrhythmic mortality during follow-up. Home Monitoring was used in all patients to continuously monitor the system integrity, device programming and performance.ConclusionThe NO Regular Defibrillation testing In Cardioverter Defibrillator Implantation (NORDIC ICD) trial is one of two large prospective randomized trials assessing the effect of DF testing omission during ICD implantation.Clinicaltrialsgov IdentifierNCT01282918.Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2014. For permissions please email: journals.permissions@oup.com.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.