J Interv Card Electrophysiol · Jan 2011
Particle formation and risk of embolization during transseptal catheterization: comparison of standard transseptal needles and a new radiofrequency transseptal needle.
- Gregory K Feld, Jay Tiongson, and Ganiyu Oshodi.
- Department of Medicine, Division of Cardiology, Cardiac Electrophysiology Program UCSD Medical Center, 4169 Front St., San Diego, CA 92103-8649, USA. gfeld@ucsd.edu
- J Interv Card Electrophysiol. 2011 Jan 1;30(1):31-6.
ObjectiveAnecdotally, the Brockenbrough transseptal needle generates plastic particles through a process of skiving (shaving off particles), when advanced through the dilator and sheath. This study was performed to assess particle creation by the Brockenbrough needle during transseptal catheterization. We explore strategies that may reduce this phenomenon, including use of the Brockenbrough stylet and a radiofrequency transseptal needle.MethodIn vitro simulations of transseptal catheterization were performed using Brockenbrough transseptal needles and a new radiofrequency transseptal needle. Particles that were created during advancement of transseptal needles through the sheath and dilator were collected and analyzed. Particles in the visible range of 50 μm to 4 mm were identified using a light microscope, whereas particles in the sub-visible, yet clinically relevant range of 10 to 50 μm, were counted using a light obscuration method.ResultsAll simulated procedures using the Brockenbrough transseptal needles, with or without a stylet, generated visible particles. Simulated procedures with the radiofrequency transseptal needle generated no visible particles. A greater number of sub-visible particles were generated with the standard Brockenbrough transseptal needle (BKR-1) without stylet compared with the standard Brockenbrough needle (BRK-1) with stylet, the Brockenbrough extra sharp (BRK-1XS) needle with or without stylet, and the radiofrequency needle (NRG C1).ConclusionClinically relevant particles, both visible and sub-visible, with the potential for causing embolic complications, are generated by the BRK-1 needle without stylet. Use of a stylet in the BRK-1 needle, or the BRK-1XS needle with or without stylet, appears to reduce the size and amount of particles created. The NRG C1 needle appears to eliminate visible particles and is comparable to the BRK-1 with stylet and the BRK-1XS with or without stylet in generation of sub-visible particles. Important steps can be taken to minimize the creation of particles during the advancement of the BRK-1 through the transseptal sheath and dilator.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.