• J Cardiothorac Surg · Jan 2013

    Case Reports

    Thrombosis during off pump LVAD placement in a patient with heparin induced thrombocytopenia using bivalirudin.

    • Hamdy Awad, Richard Bryant, Obaid Malik, Galina Dimitrova, and Chittoor Bhaskar Sai-Sudhakar.
    • J Cardiothorac Surg. 2013 Jan 1;8:115.

    AbstractHere we present our attempt at off pump HeartMate II left ventricular assist device (LVAD) implantation using the anticoagulant bivalirudin in a patient with heparin induced thrombocytopenia, which resulted in thrombosis within the LVAD device. This required that our procedure be converted to on pump, and a new HeartMate II LVAD device to be implanted. In our view, this thrombotic event may have been caused by a number of factors that include bivalirudin's (1) short half-life of about 20 minutes, (2) decreased activity with blood stasis, (3) inability to prevent clot propagation, and (4) uncertainty with real-time monitoring of therapeutic levels. To prevent future thrombotic events, it may be beneficial to immediately deair the LVAD device prior to the coring of the left ventricle during off pump LVAD placement. In addition, a solution other than blood may be used for priming. If blood is used for priming of the LVAD device, the duration of blood stasis should not exceed 20 minutes when bivalirudin is being used for anticoagulation. Furthermore, this case emphasizes the importance of having a backup LVAD device available and ready to use during surgical procedures.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…