• Anesthesia and analgesia · May 2013

    Multicenter Study

    A multicenter evaluation of a compact, sterile, single-use pressure transducer for central venous catheter placement.

    • Andrew Bowdle, Kei Togashi, Koichiro Nandate, Carli Hoaglan, and Benjamin Sherman.
    • Department of Anesthesiology, University of Washington, Seattle, Washington 98195, USA. ktogashi@u.washington.ed
    • Anesth. Analg.. 2013 May 1;116(5):1018-23.

    BackgroundInadvertent arterial placement of a large-bore catheter during attempted placement of a central venous catheter (CVC) occurs at a rate of 0.1% to 1.0% and may result in hemorrhage, pseudoaneurysm, stroke, or death. Ultrasound guidance or observation of color and pulsatility of blood are not reliable methods for avoiding this serious complication. Measurement of pressure in the needle or short plastic catheter before insertion of the guidewire has been shown to be highly reliable; however, traditional pressure measurement methodology is cumbersome. Recently a compact, sterile, single-use pressure transducer with an integrated digital display has become available. In this study, we evaluated the performance of this new device (Compass® Vascular Access).MethodsIn this prospective, observational study at 4 academic medical centers 298 CVCs were placed. Pressure was measured using the Compass transducer before and after guidewire insertion. Other details of the procedure were at the discretion of the clinician. Data describing the CVC placement and any complications were collected.ResultsTrainees placed 279 of 298 CVCs. Ultrasound guidance was used for 286 of 298 CVCs. Seven of the CVC placements occurred in the intensive care unit, with the balance occurring in the operating room. Ten of the CVCs were placed in a subclavian vein, with the balance being internal jugular vein. Two hundred seventy-four of 298 CVCs were placed on the right side. Venous pressure measured before and after guidewire insertion was 7.2 ± 4.3 (SD) and 6.5 ± 4.3 (SD) mm Hg respectively (P = 0.03). The satisfaction score recorded by the physician performing the procedure was 8.0 ± 2.1 (SD; visual analog scale 1-10, 10 being most satisfying). There were 5 inadvertent arterial punctures (1.7%). Ultrasound guidance was used in all 5 cases of arterial puncture. All of the arterial punctures were recognized before guidewire insertion by measurement of arterial pressure with the Compass transducer. No guidewires or CVC catheters were placed in arteries.ConclusionThe Compass pressure transducer for CVC placement performed as intended in 298 cases from 4 academic medical centers. There were 5 inadvertent arterial punctures despite the use of ultrasound guidance, all of which were correctly identified by pressure measurement using the Compass. The device was easily used by trainees, and users expressed a positive level of satisfaction.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.