• Critical care medicine · Feb 2008

    Controlled Clinical Trial

    Detecting ineffective triggering in the expiratory phase in mechanically ventilated patients based on airway flow and pressure deflection: feasibility of using a computer algorithm.

    • Chang-Wen Chen, Wei-Chieh Lin, Chih-Hsin Hsu, Kuo-Sheng Cheng, and Chien-Shun Lo.
    • Medical Intensive Care Unit, Department of Internal Medicine, National Cheng Kung University Medical Center, Tainan, Taiwan.
    • Crit. Care Med. 2008 Feb 1;36(2):455-61.

    ObjectiveIneffective triggering (IT) is the most common manifestation of patient-ventilator asynchrony in mechanically ventilated patients. IT in the expiratory phase (ITE) accounts for the majority of IT and is associated with characteristic features of flow and airway pressure deflection, caused by ineffective effort from the patient. The purpose of this study was to quantify the characteristics of flow and airway pressure deflections of ITE and, using a computerized algorithm, to evaluate their usefulness in the detection of ITEs.DesignProspective, clinical study.SettingMedical intensive care unit in a 1,000-bed university hospital.PatientsA total of 14 mechanically ventilated adult patients with patient-ventilator asynchrony.InterventionsNone.Measurements And Main ResultsWe analyzed 5,899 breaths and found that 1,831 were ITEs. The average values for maximum flow deflection (F(def)) and maximum airway pressure deflection (P(def)) in ITEs were 13.94 +/- 8.0 L/min and 1.91 +/- 0.97 cm H2O. With a starting value of 0.1 L/min for F(def) and 0.01 cm H2O for P(def), the area under the receiver operating characteristics curve of F(def) and P(def) for the detection of ITEs was 0.98 and 0.97, respectively. Sensitivity and specificity for the detection of ITEs were 91.5% and 96.2% for F(def), respectively, for a cutoff value of 5.45 L/min, and 93.3% and 92.9% for Pdef, for a cutoff value of 0.45 cm H2O.ConclusionWe conclude that accurately detecting and quantifying ITEs is feasible using a computerized algorithm based on F(def) and P(def). Such a computerized estimation of patient-ventilator interaction might be helpful for adjusting ventilator settings in an intensive care unit.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?

    User can't be blank.

    Content can't be blank.

    Content is too short (minimum is 15 characters).

    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…