• Int J Spine Surg · Jan 2012

    Comparison of in vivo and simulator-retrieved metal-on-metal cervical disc replacements.

    • Steven M Kurtz, Lauren Ciccarelli, Megan L Harper, Ryan Siskey, Jacob Shorez, and Frank W Chan.
    • Exponent, Inc., Philadelphia, PA ; Drexel University, Philadelphia, PA.
    • Int J Spine Surg. 2012 Jan 1;6:145-56.

    BackgroundCervical disc arthroplasty is regarded as a promising treatment for myelopathy and radiculopathy as an alternative to cervical spine fusion. On the basis of 2-year clinical data for the PRESTIGE(®) Cervical Disc (Medtronic, Memphis, Tennessee), the Food and Drug Administration recommended conditional approval in September 2006 and final approval in July 2007; however, relatively little is known about its wear and damage modes in vivo. The main objective was to analyze the tribological findings of the PRESTIGE(®) Cervical Disc. This study characterized the in vivo wear patterns of retrieved cervical discs and tested the hypothesis that the total disc replacements exhibited similar surface morphology and wear patterns in vitro as in vivo.MethodsTen explanted total disc replacements (PRESTIGE(®), PRESTIGE(®) I, and PRESTIGE(®) II) from 10 patients retrieved after a mean of 1.8 years (range, 0.3-4.1 years) were analyzed. Wear testing included coupled lateral bending ( ±4.7°) and axial rotation ( ±3.8°) with a 49 N axial load for 5 million cycles followed by 10 million cycles of flexion-extension ( ±9.7°) with 148 N. Implant surfaces were characterized by the use of white-light interferometry, scanning electron microscopy, and energy dispersive spectroscopy.ResultsThe explants generally exhibited a slightly discolored, elliptic wear region of varying dimension centered in the bearing center, with the long axis oriented in the medial-lateral direction. Abrasive wear was the dominant in vivo wear mechanism, with microscopic scratches generally oriented in the medial-lateral direction. Wear testing resulted in severe abrasive wear in a curvilinear fashion oriented primarily in the medial-lateral direction. All retrievals showed evidence of an abrasive wear mechanism.ConclusionsThis study documented important similarity between the wear mechanisms of components tested in vitro and explanted PRESTIGE(®) Cervical Discs; however, the severity of wear was much greater during the in vitro test compared with the retrievals.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.