• Neurosurgery · Dec 2008

    The new Licox combined brain tissue oxygen and brain temperature monitor: assessment of in vitro accuracy and clinical experience in severe traumatic brain injury.

    • Campbell Stewart, Iain Haitsma, Zsolt Zador, J Claude Hemphill, Diane Morabito, Geoffrey Manley, and Guy Rosenthal.
    • Department of Surgery, San Francisco General Hospital, University of California, San Francisco, San Francisco, California 94110, USA.
    • Neurosurgery. 2008 Dec 1;63(6):1159-64; discussion 1164-5.

    ObjectiveMonitoring of brain tissue oxygen tension is increasingly being used to monitor patients after severe traumatic brain injury and to guide therapies aimed at maintaining brain tissue oxygen tension above threshold levels. The new Licox PMO combined oxygen and temperature catheter (Integra LifeSciences, Plainsboro, NJ) combines measurements of oxygen tension and temperature in a single probe inserted through a bolt mechanism. In this study, we sought to evaluate the accuracy of the new Licox PMO probe under controlled laboratory conditions and to assess the accuracy of oxygen tension and temperature measurements and the new automated card calibration system. We also describe our clinical experience with the Licox PMO probe.MethodsOxygen tension was measured in a 2-chambered apparatus at different oxygen tensions and temperatures. The new card calibration system was compared with a manually calibrated system. Rates of hematoma, infection, and dislodgement in our clinical experience were recorded.ResultsThe new Licox PMO probe accurately measures oxygen tension over a wide range of oxygen concentrations and physiological temperatures, but it does have a small tendency to underestimate oxygen tension (mean error, -3.8 +/- 3.5%) that is more pronounced between the temperatures of 33 and 39 degrees C. The thermistor of the PMO probe also has a tendency to underestimate temperature when compared with a resistance thermometer (mean error, -0.67 +/- 0.22 degrees C). The card calibration system was also found to introduce some variability in measurements of oxygen tension when compared with a manually calibrated system. Clinical experience with the new probe indicates good placement within the white matter using the improved bolt system and low rates of hematoma (2.9%), infection (0%), and dislodgement (5.9%).ConclusionThe new Licox PMO probe is accurate but has a small, consistent tendency to under-read oxygen tension that is more pronounced at higher temperatures. The probe tends to under-read temperature by 0.5 to 0.8 degrees C across temperatures, suggesting that caution should be used when brain temperature is measured with the Licox PMO probe and used to guide temperature-directed treatment strategies. The Licox PMO probe improves upon previous models in allowing consistent and accurate placement in the white matter and obviating the need for placement of 2 separate probes to measure oxygen tension and temperature.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…