- Chun-Yao Huang, Din-E Shan, Chao-Hung Lai, Man-Cai Fong, Po-Shun Huang, Hsien-Hao Huang, Shing-Jong Lin, and Chern-En Chiang.
- Int. J. Cardiol. 2006 Jul 28;111(1):163-5.
AbstractTremor-induced electrocardiographic artifacts could be misdiagnosed as ventricular tachycardia (VT). However, there has been no electrocardiographic algorithm effectively differentiating pseudo-VT. In this study, we used 3 electrocardiographic "signs": "Sinus" sign, "Spike" sign, and "Notch" sign, and created an electrocardiographic algorithm. The algorithm was prospectively tested in 98 electrocardiographs (37 tremor-induced pseudo-VT and 61 true VT) Thirty-six out of 37 (97.3%) tremor-induced pseudo-VTs could be accurately diagnosed. In conclusion, this is the first study to systemically analyze the tremor-induced pseudo-VT. Our new electrocardiographic algorithm provides a useful tool for a rapid and accurate diagnosis.