• Anesthesia and analgesia · Jul 1993

    Laboratory evaluation of the Baxter Patient-Controlled Analgesia Infusor System: a disposable patient-controlled analgesia device.

    • N A Mackey, A H Ilsley, H Owen, and J L Plummer.
    • Department of Anesthesia and Intensive Care, Flinders Medical Centre, Adelaide, South Australia.
    • Anesth. Analg. 1993 Jul 1;77(1):117-20.

    AbstractThe purpose of this study was to determine the accuracy of bolus-demand volumes obtained from the Baxter Patient-Controlled Analgesia Infusor System, a disposable patient-controlled analgesia device. An automated system was used to make demands in a pattern designed to simulate those made by a patient in the first 24 postoperative hours. Following this, a pattern of varying short interdemand intervals was used to evaluate performance of the device when demands were made near the nominal lockout interval of 6 min. Finally, a prolonged array of short interdemand intervals was used to evaluate performance as the reservoir of the devices became exhausted. Ten devices were tested. One device failed; after 70 min of testing the reservoir burst. The mean delivery of the remaining nine devices for the first 24 h of testing was 92.6% of nominal (SD = 3.9%). Evaluation near the nominal lockout interval revealed a departure of delivered volume from nominal (worst = 73.9% at an interdemand interval of 6 min). As the reservoir became exhausted, delivery increased to a maximum value, ranging from 118% to 137% of nominal, before falling to zero. In general, the devices performed consistently and the departures from nominal delivery were not considered to be of clinical significance.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…