Thermal softening of tracheal tubes: an unrecognized hazard of the Bair Hugger active patient warming system.
- J L Ayala and A Coe.
- Department of Anaesthesia, Scunthorpe Hospital NHS Trust, Cliff Gardens.
- Br J Anaesth. 1997 Oct 1;79(4):543-5.
AbstractThe Bair Hugger system is a new and highly effective active patient warming system which produces a layer of warm air between the patient and the warming system. We report an instance of marked softening and distortion of a polyvinyl chloride tracheal tube caused by this layer. We also present laboratory data indicating that this is a likely problem under routine theatre conditions, with suggestions for prevention.