• Critical care medicine · May 2002

    Significant increase of activated partial thromboplastin time by heparinization of the radial artery catheter flush solution with a closed arterial catheter system.

    • Eric A J Hoste, Nathalie R A Roels, Johan M A Decruyenaere, and Francis A Colardyn.
    • Intensive Care Unit, Ghent University Hospital, Ghent, Belgium. erik.hoste@rug.ac.be
    • Crit. Care Med. 2002 May 1;30(5):1030-4.

    ObjectiveEvaluate whether the use of a heparinized flush for an arterial catheter with a closed-loop blood sampling system leads to erroneous coagulation studies.DesignProspective study.SettingTwenty-two-bed surgical adult intensive care unit in a university hospital.PatientsSixty patients, 30 in each phase of the study.InterventionsPhase 1: coagulation studies on a blood sample from a venous puncture and an arterial blood sample from an arterial catheter with heparinized flush. Phase 2: the same protocol but the arterial catheter was flushed with saline.MeasurementsActivated partial thromboplastin time, fibrinogen, percentage of prothrombin time, and international normalized ratio of prothrombin time on the venous and arterial samples.Main ResultsActivated partial thromboplastin time in the arterial blood samples taken from an arterial catheter with heparinized flush was significantly prolonged compared with the venous controls. This was not the case in blood samples from an arterial catheter with saline flush. The magnitude of this difference found by the Bland and Altman analysis was clinically significant. In 23.3% of the patients in phase 1, there was a difference of >50% between the arterial and venous sampled activated partial thromboplastin time, compared with 0% of the patients in phase 2 (p =.018). There was no difference between the venous and the arterial blood samples for the non-heparin-sensitive coagulation studies in both phases of the study. An explanation for these findings could be that there was release of heparin bound to the arterial catheter into the blood sample.ConclusionA heparinized flush solution for the arterial catheter, when used together with a closed-loop blood sampling system, leads to erroneous results of heparin-sensitive coagulation studies. Heparin-sensitive coagulation studies, therefore, should not be analyzed on blood samples from such a system if a heparinized flush solution is used.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…