    Reliability and accuracy of the thoracic impedance signal for measuring cardiopulmonary resuscitation quality metrics.

    • Erik Alonso, Jesús Ruiz, Elisabete Aramendi, Digna González-Otero, Sofía Ruiz de Gauna, Unai Ayala, James K Russell, and Mohamud Daya.
    • Communications Engineering Department, University of the Basque Country UPV/EHU, Alameda Urquijo S/N, 48013 Bilbao, Spain. Electronic address: erik_alonso@ehu.es.
    • Resuscitation. 2015 Mar 1;88:28-34.

    AimTo determine the accuracy and reliability of the thoracic impedance (TI) signal to assess cardiopulmonary resuscitation (CPR) quality metrics.MethodsA dataset of 63 out-of-hospital cardiac arrest episodes containing the compression depth (CD), capnography and TI signals was used. We developed a chest compression (CC) and ventilation detector based on the TI signal. TI shows fluctuations due to CCs and ventilations. A decision algorithm classified the local maxima as CCs or ventilations. Seven CPR quality metrics were computed: mean CC-rate, fraction of minutes with inadequate CC-rate, chest compression fraction, mean ventilation rate, fraction of minutes with hyperventilation, instantaneous CC-rate and instantaneous ventilation rate. The CD and capnography signals were accepted as the gold standard for CC and ventilation detection respectively. The accuracy of the detector was evaluated in terms of sensitivity and positive predictive value (PPV). Distributions for each metric computed from the TI and from the gold standard were calculated and tested for normality using one sample Kolmogorov-Smirnov test. For normal and not normal distributions, two sample t-test and Mann-Whitney U test respectively were applied to test for equal means and medians respectively. Bland-Altman plots were represented for each metric to analyze the level of agreement between values obtained from the TI and gold standard.ResultsThe CC/ventilation detector had a median sensitivity/PPV of 97.2%/97.7% for CCs and 92.2%/81.0% for ventilations respectively. Distributions for all the metrics showed equal means or medians, and agreements >95% between metrics and gold standard was achieved for most of the episodes in the test set, except for the instantaneous ventilation rate.ConclusionWith our data, the TI can be reliably used to measure all the CPR quality metrics proposed in this study, except for the instantaneous ventilation rate.Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,706,642 articles already indexed!

We guarantee your privacy. Your email address will not be shared.