• Critical care medicine · Jun 1996

    Comparative Study

    Hand-held blood gas analyzer is accurate in the critical care setting.

    • G P Zaloga, P R Roberts, K Black, J T Santamauro, E Klase, and M Suleiman.
    • Section on Critical Care, Department of Anesthesia, Bowman Gray School of Medicine of Wake Forest University, Winston-Salem, NC 27157-1009, USA.
    • Crit. Care Med. 1996 Jun 1; 24 (6): 957-62.

    ObjectiveTo determine the accuracy of a new, portable battery-powered blood gas analyzer when used by nonlaboratory-trained clinicians in the critical care setting.DesignProspective analysis of blood samples from critically ill patients.SettingLarge tertiary critical care unit.PatientsHeterogeneous group of medical and surgical critically ill patients.InterventionsNone.Measurements And Main ResultsTwo hundred thirty-nine split blood samples from intensive care patients were analyzed by clinicians in the critical care environment using a new, portable, battery-powered blood gas analyzer (Immediate Response Mobile Analyzer [IRMA], Diametrics Medical, St. Paul, MN). Near-patient measurements were compared with measurements obtained by laboratory technologists using an IL-1312 blood gas analyzer (Instrumentation Laboratories, Lexington, MA) in an established near-patient critical care laboratory. Precision and coefficients of variation were also determined using repeated testing of quality control samples at three levels of pH, PO2, and PCO2. There was good agreement between IRMA determinations and the laboratory. Correlation coefficients ranged from 0.96 to 0.99. Bias and precision (+/-2 SD), respectively, were 0.02 and 0.036 units for pH, -0.3 torr (1.8 kPa) (-0.04 kPa) and 7.2 torr (0.96 kPa) for PCO2 and -3.9 torr (0.52 kPa) and 13.8 torr (1.8 kPa) for PO2. Precision on repeated testing of quality control samples ranged from 0.022 to 0.04 units for a pH of 7.2 to 7.6, 1.2 to 4.6 torr (0.16 to 0.61 kPa) for a PCO2 of 20 to 60 torr (2.7 to 8 kPa), and 3.0 to 7.4 torr (0.40 to 0.99 kPa) for a PO2 of 70 to 160 torr (9.3 to 21.3 kPa). Coefficients of variation ranged from 0.15% to 0.28% for a pH of 7.2 to 7.6, 2.0% to 3.7% for a PCO2 of 20 to 60 torr (2.7 to 8.0 kPa), and 1.7% to 3.6% for a PO2 of 70 to 160 torr (9.3 to 21.3 kPa). Mean turnaround time was 16.5 +/-10.1 mins for the near-patient laboratory and 2+/-0.5 mins for IRMA.ConclusionsIRMA is accurate and reproducible when used in the clinical setting by nonlaboratory-trained individuals. Nonlaboratory-trained individuals can obtain laboratory results in the critical care setting comparable with the results obtained by trained laboratory technologists. Bedside laboratory testing decreases turnaround time compared with a near-patient laboratory.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…