- M W Allan, W M Gray, and A J Asbury.
- Department of Anaesthetics, Western Infirmary, Glasgow.
- Br J Anaesth. 1988 Mar 1; 60 (4): 413-8.
AbstractThe dynamic performance of arterial manometers depends upon the values of the natural frequency fo and the damping factor beta. Satisfactory reproduction of all arterial waveforms requires that fo should be greater than about 13 Hz; for an fo of 13 Hz, beta should have a value of 0.5, but, as fo increases, an increasing range of beta values is permissible. We measured fo and beta for two pressure transducers (Gould P23 Statham and Gould Disposable) in combination with Portex or Gould catheter tubing of lengths varying from 100 cm to 300 cm. All combinations were considerably underdamped, with the best having an fo of 30 Hz and a beta of 0.1. We also investigated the performance of a commercial device, the Accudynamic, designed to allow beta to be increased without affecting fo. We found that the Accudynamic worked well, allowing acceptable performance to be obtained from the Gould Disposable transducer with catheters up to 200 cm in length.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.