• J Trauma · May 2009

    Micromotion at the fracture site after tibial nailing with four unreamed small-diameter nails--a biomechanical study using a distal tibia fracture model.

    • Michael Schüller, Patrick Weninger, Elmar Tschegg, Michael Jamek, Heinz Redl, and Stefanie Stanzl-Tschegg.
    • Institute for Material Sciences, Technical University, Vienna, Austria.
    • J Trauma. 2009 May 1; 66 (5): 1391-7.

    BackgroundIntramedullary nailing is the treatment of choice in tibia fractures allowing for closed fracture reduction and internal fixation. Small-diameter nails that preserve the endosteal blood supply act as load-sharing devices after proximal and distal locking. Despite fracture healing is influenced by movements at the fracture gap, no data are available reporting on the micromovements at the fracture site if small-diameter nails were used.MethodsUsing a Sawbone distal tibia fracture model, we assessed offset, elastic, plastic, permanent, and overall deformation at the fracture site for four small-diameter tibia nails (Expert, Synthes, Saluburg, Austria; Connex, ITS Spectromed, Lassnitzhöhe, Austria; Versanail, DePuy, Vienna, Austria; T2, Stryker, Vienna, Austria) after mechanical testing with a servohydraulic material testing machine. Cyclic loading was performed with a sinusoidal load of 700 N (+/-600) for 40,000 cycles representing 6 weeks of full weight bearing.ResultsOffset deformation was significantly higher for the Connex nail when compared with other nails (p < 0.001). Regarding elastic deformation, no significant difference was recorded between the implants. Plastic deformation was significantly lower if the Connex nail was used (0.134 [+/-0.053] mm; p < 0.001). Elastic deformation did not exceed 0.7 mm and plastic deformation did not exceed 0.4 mm. Regarding permanent and overall deformation, no significant difference between the implants was recorded.ConclusionsConsiderable deformation at the fracture gap can be assumed even after partial weight bearing with 100 N. Despite comparable material properties, differences in axial micromotion were recorded among the nail types used in this series. The number of distal locking screws (three or four) did not substantially influence the axial movements at the fracture gap.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…