• N. Engl. J. Med. · Feb 2017

    Multicenter Study

    Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator.

    • Robert J Russo, Heather S Costa, Patricia D Silva, Jeffrey L Anderson, Aysha Arshad, Robert W W Biederman, Noel G Boyle, Jennifer V Frabizzio, Ulrika Birgersdotter-Green, Steven L Higgins, Rachel Lampert, Christian E Machado, Edward T Martin, Andrew L Rivard, Jason C Rubenstein, Raymond H M Schaerf, Jennifer D Schwartz, Dipan J Shah, Gery F Tomassoni, Gail T Tominaga, Allison E Tonkin, Seth Uretsky, and Steven D Wolff.
    • From the Scripps Research Institute (R.J.R.), the La Jolla Cardiovascular Research Institute (R.J.R., P.D.S.), University of California, San Diego (U.B.-G.), and Scripps Memorial Hospital (S.L.H., G.T.T.), La Jolla, the University of California, Los Angeles, Los Angeles (N.G.B.), and Providence St. Joseph Medical Center, Burbank (R.H.M.S.) - all in California; the Department of Entomology, University of Arizona, Tucson (H.S.C.); Intermountain Medical Center, Salt Lake City (J.L.A., A.E.T.); Inova Heart and Vascular Institute, Falls Church, VA (A.A.); Allegheny General Hospital, Pittsburgh (R.W.W.B.), and Abington Memorial Hospital, Abington (J.V.F.) - both in Pennsylvania; Yale University School of Medicine, New Haven, CT (R.L.); Providence Heart Institute, Southfield, MI (C.E.M.); Oklahoma Heart Institute, Tulsa (E.T.M.); University of Mississippi Medical Center, Jackson (A.L.R.); Medical College of Wisconsin, Milwaukee (J.C.R.); Bassett Medical Center, Cooperstown (J.D.S.), and Advanced Cardiovascular Imaging, Carnegie Hill Radiology, New York (S.U., S.D.W.) - both in New York; Methodist DeBakey Heart and Vascular Center, Houston (D.J.S.); and Baptist Health, Lexington, KY (G.F.T.).
    • N. Engl. J. Med. 2017 Feb 23; 376 (8): 755-764.

    AbstractBackground The presence of a cardiovascular implantable electronic device has long been a contraindication for the performance of magnetic resonance imaging (MRI). We established a prospective registry to determine the risks associated with MRI at a magnetic field strength of 1.5 tesla for patients who had a pacemaker or implantable cardioverter-defibrillator (ICD) that was "non-MRI-conditional" (i.e., not approved by the Food and Drug Administration for MRI scanning). Methods Patients in the registry were referred for clinically indicated nonthoracic MRI at a field strength of 1.5 tesla. Devices were interrogated before and after MRI with the use of a standardized protocol and were appropriately reprogrammed before the scanning. The primary end points were death, generator or lead failure, induced arrhythmia, loss of capture, or electrical reset during the scanning. The secondary end points were changes in device settings. Results MRI was performed in 1000 cases in which patients had a pacemaker and in 500 cases in which patients had an ICD. No deaths, lead failures, losses of capture, or ventricular arrhythmias occurred during MRI. One ICD generator could not be interrogated after MRI and required immediate replacement; the device had not been appropriately programmed per protocol before the MRI. We observed six cases of self-terminating atrial fibrillation or flutter and six cases of partial electrical reset. Changes in lead impedance, pacing threshold, battery voltage, and P-wave and R-wave amplitude exceeded prespecified thresholds in a small number of cases. Repeat MRI was not associated with an increase in adverse events. Conclusions In this study, device or lead failure did not occur in any patient with a non-MRI-conditional pacemaker or ICD who underwent clinically indicated nonthoracic MRI at 1.5 tesla, was appropriately screened, and had the device reprogrammed in accordance with the prespecified protocol. (Funded by St. Jude Medical and others; MagnaSafe ClinicalTrials.gov number, NCT00907361 .).

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…