Randomized Controlled Trial
Radiation dose reduction in thoracic and lumbar spine instrumentation using navigation based on an intraoperative cone beam CT imaging system: a prospective randomized clinical trial.
- Nathalie Pireau, Virginie Cordemans, Xavier Banse, Nadia Irda, Sébastien Lichtherte, and Ludovic Kaminski.
- Service d'orthopédie et de traumatologie de l'appareil locomoteur, Cliniques Universitaires Saint-Luc, Avenue Hippocrate 10, B-1200, Brussels, Belgium.
- Eur Spine J. 2017 Nov 1; 26 (11): 2818-2827.
PurposeSpine surgery still remains a challenge for every spine surgeon, aware of the potential serious outcomes of misplaced instrumentation. Though many studies have highlighted that using intraoperative cone beam CT imaging and navigation systems provides higher accuracy than conventional freehand methods for placement of pedicle screws in spine surgery, few studies are concerned about how to reduce radiation exposure for patients with the use of such technology. One of the main focuses of this study is based on the ALARA principle (as low as reasonably achievable).MethodA prospective randomized trial was conducted in the hybrid operating room between December 2015 and December 2016, including 50 patients operated on for posterior instrumented thoracic and/or lumbar spinal fusion. Patients were randomized to intraoperative 3D acquisition high-dose (standard dose) or low-dose protocol, and a total of 216 pedicle screws were analyzed in terms of screw position. Two different methods were used to measure ionizing radiation: the total skin dose (derived from the dose-area product) and the radiation dose evaluated by thermoluminescent dosimeters on the surgical field.ResultsAccording to Gertzbein and Heary classifications, low-dose protocol provided a significant higher accuracy of pedicle screw placement than the high-dose protocol (96.1 versus 92%, respectively). Seven screws (3.2%), all implanted with the high-dose protocol, needed to be revised intraoperatively. The use of low-dose acquisition protocols reduced patient exposure by a factor of five.ConclusionThis study emphasizes the paramount importance of using low-dose protocols for intraoperative cone beam CT imaging coupled with the navigation system, as it at least does not affect the accuracy of pedicle screw placement and irradiates drastically less.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.