• J. Thorac. Cardiovasc. Surg. · Feb 2019

    Stent and leaflet stresses in 26-mm, third-generation, balloon-expandable transcatheter aortic valve.

    • Yue Xuan, Danny Dvir, Zhongjie Wang, Trek Mizoguchi, Jian Ye, Julius M Guccione, Liang Ge, and Elaine E Tseng.
    • Department of Surgery, University of California San Francisco and San Francisco VA Medical Centers, San Francisco, Calif.
    • J. Thorac. Cardiovasc. Surg. 2019 Feb 1; 157 (2): 528-536.

    ObjectivesTranscatheter aortic valve replacement has proven successful in treating intermediate-risk, high-risk, and inoperable patients with severe aortic stenosis. Third-generation, balloon-expandable transcatheter aortic valves were developed with an outer sealing skirt to reduce paravalvular leakage. As transcatheter aortic valve replacement use expands, long-term durability questions remain. Valve design influences durability, where regions of increased leaflet stress are vulnerable to early degeneration. However, third-generation transcatheter aortic valve stresses are unknown. Our goals were to determine the stent and leaflet stresses of third-generation, balloon-expandable transcatheter aortic valves.MethodsThe commercial 26-mm Edwards SAPIEN 3 valve (Edwards Lifesciences, Inc, Irvine, Calif) underwent high-resolution micro-computed tomography scanning to develop a precise 3-dimensional geometric mesh of the stent and valve. Leaflet material properties were obtained from surgical bioprostheses, and stent material properties were based on cobalt-chromium. Simulations of systemic pressure loading were performed, and stress was calculated using finite element analyses.ResultsAt diastole, maximum and minimum principal stresses on transcatheter aortic valve leaflets were 2.7 MPa and -0.47 MPa, respectively. Peak leaflet stresses were observed at upper leaflet commissures, at their connection to the stent. Maximum and minimum principal stresses for the stent were 38.2 MPa and -44.4 MPa, respectively, at 80 mm Hg and were located just below the commissural stent.ConclusionsStress analysis of the 26-mm SAPIEN 3 valve using exact geometry from high-resolution scans demonstrated that peak stresses for both transcatheter aortic valve stent and leaflets were present at commissural tips where leaflets were attached. These regions would be most likely to initiate degeneration. The Dacron skirt had minimal effect on stresses on leaflets and stent.Published by Elsevier Inc.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.