• J Clin Monit Comput · Feb 2019

    Observational Study

    Continuous noninvasive arterial blood pressure monitoring using the vascular unloading technology during complex gastrointestinal endoscopy: a prospective observational study.

    • Julia Y Nicklas, David Beckmann, Jasper Killat, Martin Petzoldt, Daniel A Reuter, Thomas Rösch, and Bernd Saugel.
    • Department of Anesthesiology, Center of Anesthesiology and Intensive Care Medicine, University Medical Center Hamburg-Eppendorf, Martinistrasse 52, 20246, Hamburg, Germany.
    • J Clin Monit Comput. 2019 Feb 1; 33 (1): 25-30.

    AbstractThe innovative vascular unloading technology (VUT) allows continuous noninvasive arterial blood pressure (AP) monitoring. We aimed to investigate whether the VUT enables AP changes to be detected earlier compared with intermittent AP monitoring in patients undergoing gastrointestinal endoscopy. In this prospective observational study, we recorded continuous AP measurements with the VUT (CNAP system; CNSystems Medizintechnik AG, Graz, Austria) and intermittent AP measurements with upper arm cuff oscillometry in 90 patients undergoing complex gastrointestinal endoscopy (Department of Interventional Endoscopy at the University Medical Center Hamburg-Eppendorf, Hamburg, Germany). A "hypotensive phase" was defined as a time period of at least 30 s during which ≥ 50% of the VUT-AP values were in a predefined range of hypotension, i.e., AP value a) ≥ 10% below the last oscillometric value and b) ≤ 65 mmHg for mean AP or ≤ 90 mmHg for systolic AP. In the 5-min-interval between two oscillometric measurements, one or more hypotensive phases were detected in 26 patients (29%) for mean AP and in 27 patients (30%) for systolic AP. Hypotensive phases had a mean duration of 195 ± 99 s for mean AP and 197 ± 97 s for systolic AP with a mean procedure duration of 36 (± 21) min. Continuous noninvasive AP monitoring using the VUT enables hypotensive phases to be detected earlier compared with intermittent AP monitoring during complex gastrointestinal endoscopy. These hypotensive phases may be missed or only belatedly recognized with intermittent AP monitoring. Continuous noninvasive AP measurement facilitates detecting hemodynamic instability more rapidly and therefore may improve patient safety.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…