Eur. J. Heart Fail. · Oct 2014
New evidence for favourable effects on haemodynamics and ventricular performance after Parachute(®) implantation in humans.
- Tobias Schmidt, Christian Frerker, Thomas Thielsen, Inge Dotz, Peter Wohlmuth, Karl-Heinz Kuck, and Ulrich Schäfer.
- ASKLEPIOS Klinik St. Georg Hospital, Division of Cardiology, Lohmühlenstrasse 5, 20099, Hamburg, Germany.
- Eur. J. Heart Fail. 2014 Oct 1; 16 (10): 1112-9.
AimsThe Parachute(®) Ventricular Partitioning Device offers an additional strategy for heart failure patients with exclusion of the infarcted wall to decrease left ventricular volumes, myocardial work, and wall stress. The aim of the present study was to evaluate if Parachute implantation might influence acute haemodynamic and functional performance in patients with left ventricular aneurysm after anteroapical infarction.Methods And ResultsSixteen patients underwent a Parachute device implantation. Invasive right and left heart haemodynamic assessments as well as left ventricular analysis for evaluating left ventricle end-diastolic and end-systolic volumes, and regional ventricular function were analysed. After implantation a significant increase in stroke volume (+25.4%, P = 0.0005), stroke volume index (+26.5%, P = 0.0005), cardiac output (+25.8%, P < 0.0001) and cardiac index (+25.9%, P < 0.0001) was found. In addition to an increase in mean aortic (P = 0.0050) and pulmonary pressure (P = 0.0347), there were significant increases in stroke work index (P = 0.0003), left (P = 0.0015) and right (P = 0.0024) ventricular stroke work index as well as left and right cardiac work index (both P = 0.0001), while the remaining haemodynamic parameters remained unchanged. Left ventricular analysis showed an acute reduction of the left ventricular end-diastolic volume (-18.0%, P < 0.0001) and left ventricular end-systolic volume (-26.3%, P < 0.0001) and an increase in ejection fraction from 22.9 to 30.6% (+38.4%, P < 0.0001). Most interestingly, the basal wall segments displayed an increased contribution to the left ventricular ejection fraction with increased wall motion in nearly all segments (except the apex region).ConclusionThe data demonstrate the acute haemodynamic efficacy of Parachute device implantation. The implantation of the device displays immediate significant left ventricular volume reduction leading to an acute improved right and left cardiac function, proving the concept of left ventricular partitioning.© 2014 The Authors. European Journal of Heart Failure © 2014 European Society of Cardiology.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.