Brachytherapy
-
The purpose of this study was to compare an isotropic three-dimensional (3D) T2-weighted sequence sampling perfection with application-optimized contrasts by using flip angle evolution (SPACE) with an axial two-dimensional T2-weighted turbo spin echo (TSE) sequence with regard to overall image quality and the delineation of normal prostate and periprostatic anatomy for low-dose-rate prostate cancer brachytherapy planning evaluation. ⋯ Anatomic delineation of the prostatic and periprostatic anatomy provided by the 3D SPACE sequence is as robust in quality as that provided by a conventional 2D TSE sequence with superior delineation of the genitourinary diaphragm. For MRI-based brachytherapy treatment planning, the 3D SPACE sequence with subcentimeter isotropic resolution can replace the 2D TSE sequence and be incorporated into standard MRI protocols.