Research and practice in thrombosis and haemostasis
-
Res Pract Thromb Haemost · Mar 2021
Exploring the utility of a novel point-of-care whole blood thrombin generation assay following trauma: A pilot study.
Plasma thrombin generation kinetics as measured by the calibrated automated thrombogram (CAT) assay is a predictor of symptomatic venous thromboembolism after trauma. We hypothesized that data from a new prototype assay for measurement of thrombin generation kinetics in fresh whole blood (near patient testing of thrombin generation), will correlate with the standard CAT assay in the same patients, making it a potential tool in the future care of trauma patients. ⋯ In this pilot study, we have found that a novel point-of-care whole blood thrombin generation assay yields results with modest but statistically significant correlations to those of a standard plasma thrombin generation assay. This finding supports studying this device in a larger, adequately powered study.