Heart rhythm : the official journal of the Heart Rhythm Society
-
Recommendations regarding performance of magnetic resonance imaging (MRI) in non-MRI conditional pacemaker and defibrillator recipients are evolving. Previous studies have suggested low adverse event rates with MRI in nonconditional cardiac implantable electronic device (CIED) recipients, but low power limits optimal characterization of risk. ⋯ This review demonstrated low lead failure and clinical event rates in non-MRI conditional pacemaker and defibrillator recipients undergoing MRI. Observed changes were small and interstudy variance was low, suggesting that the composite event rates offer a reasonable estimate of true effect. The observed adverse events reinforce the need for ongoing vigilance and caution, particularly with older devices.