Evaluation of mobility and stability in the Discover artificial disc: an in vivo motion study using high-accuracy 3D CT data.
Artificial disc replacement (ADR) devices are unlike implants used in cervical fusion in that they are continuously exposed to stress not only within the implant site but also at their site of attachment to the adjacent vertebra. An imaging technique with higher accuracy than plain radiography and with the possibility of 3D visualization would provide more detailed information about the motion quality and stability of the implant in relation to the vertebrae. Such high-accuracy studies have previously been conducted with radiostereometric analysis (RSA), which requires implantation of tantalum markers in the adjacent vertebrae. The aim of this study was to evaluate in vivo motion and stability of implanted artificial discs. A noninvasive analysis was performed with CT, with an accuracy higher than that of plain radiographs and almost as high as RSA in cervical spine. ⋯ The majority of the artificial discs in this study showed intrinsic mobility several years after implantation and were also shown to be properly attached. Implant instability was detected in 8% of patients and, as all of these patients underwent revision surgery due to increasing neck pain, this might be a more serious problem than heterotopic bone formation.