Expert review of medical devices
-
Expert Rev Med Devices · Mar 2011
ReviewState of the art in hard-on-hard bearings: how did we get here and what have we achieved?
Total hip arthroplasty has shown excellent results in decreasing pain and improving function in patients with degenerative disease of the hip. Improvements in prosthetic materials, designs and implant fixation have now resulted in wear of the bearing surface being the limitation of this technology, and a number of hard-on-hard couples have been introduced to address this concern. The purpose of this article is to review the origins, development, survival rates and potential advantages and disadvantages of the following hard-on-hard bearings for total hip arthroplasty: metal-on-metal standard total hip arthroplasty; metal-on-metal hip resurfacing arthroplasty, ceramic-on-ceramic total hip arthroplasty; and ceramic-on-metal bearings. ⋯ Contemporary ceramic-on-metal interfaces are in the early stages of clinical use, with little data reported to date. Hard-on-hard bearings for total hip arthroplasty have improved dramatically over the past 50 years. As bearing designs continue to improve with new and modified materials and improved manufacturing techniques, it is likely that the use of hard-on-hard bearings will continue to increase, especially in young and active patients.