Clinical TrialThree-dimensional load measurements in an external fixator.
On the basis of a six-degree-of-freedom adjustable fracture reduction hexapod external fixator, a system which can be used for measuring axial and shear forces as well as torsion and bending moments in the fixator in vivo was developed. In a pilot study on 9 patients (7 fresh fractures and 2 osteotomies of the tibia), the load in the fixator during the healing process was measured after 2, 4, 8 and 12 weeks and at fixator removal. The measured values enabled both the type of fracture to be determined as well as the monitoring of the healing process. ⋯ Compared with other external fixator load measurements in vivo, the hexapod offers the advantage of being able to measure all forces and moments in the fixator separately and with a relatively simple mechanical arrangement. In our opinion, it will be possible to control fracture healing using this system, thereby minimizing radiation exposure from radiographs. Furthermore, the measurement system is a step towards the development of external fixator systems that enable automatic adjustments of the callus mechanical situation ("automatic dynamization") and inform the patients about the optimal weight bearing of their extremity ("intelligent fixator").