This article reviews the principles of pulse oximetry and assesses the accuracy of pulse oximeter measurements obtained during exercise, based on reports of 10 studies that evaluated 24 pulse oximeters. Nine of the studies used cycle exercise, and 1 study utilized treadmill running for mode of activity. Subject populations included patients with cardiovascular or pulmonary disorders, nondisabled individuals, and athletes. ⋯ Recent studies suggest the current generation of finger-probe-equipped pulse oximeters may be more accurate than ear-probe-equipped models. We recommend that clinicians carefully secure the probe; monitor signal strength; be wary of %SpO2 values of < or = 68% to 78% to avoid undetected severe hypoxemia; and be alert to whether a patient is a smoker, in which case %SpO2 will likely be overestimated. Researchers interested in using pulse oximeter estimates should validate their particular pulse oximeter(s) via arterial blood sampling.