Journal of clinical microbiology
-
J. Clin. Microbiol. · Jan 2013
Multicenter StudyClinical accuracy of a PLEX-ID flu device for simultaneous detection and identification of influenza viruses A and B.
Respiratory tract infections caused by influenza A and B viruses often present nonspecifically, and a rapid, high-throughput laboratory technique that can identify influenza viruses is clinically and epidemiologically desirable. The PLEX-ID Flu assay (Abbott Molecular Inc., Des Plaines, IL) incorporates multilocus PCR and electrospray ionization-mass spectrometry to detect and differentiate influenza A 2009 H1N1 (H1N1-p), seasonal H1N1 (H1N1-s), influenza A H3N2, and influenza B viruses in nasopharyngeal swab (NPS) specimens. The clinical performance characteristics of the PLEX-ID Flu assay in symptomatic patients were determined in this multicenter trial. ⋯ The overall agreements between the PLEX-ID and Prodesse ProFLU+/ProFAST+ assays were 97.1 to 100.0%. Bidirectional Sanger sequencing analysis revealed that 87.5% of 96 discrepant results between the PLEX-ID Flu and ProFLU+/ProFAST+ assays were found upon influenza A virus detection and H1N1-p subtyping. The PLEX-ID Flu assay demonstrated a high level of accuracy for the simultaneous detection and identification of influenza A and B viruses in patient specimens, providing a new laboratory tool for the rapid diagnosis and management of influenza A and B virus infections.