Journal of medical engineering & technology
-
Review Comparative Study
MR-compatible pumps versus manual titration of propofol for pediatric sedation.
This study aimed to review the use of two methods of propofol infusion: a metred burette system where propofol is diluted in Lactated Ringer using the rule of six [ 1 ] or an infusion pump: Medrad Continuum Magnetic resonance (MR) compatible Infusion System (Medrad Inc. Indianola, PA) and to determine the difference, if any, between the total amount of drug delivered, the impact on vital signs, sedation status and discharge time of the patients. With Institutional Review Board (IRB) approval, 140 children aged 0 to 18 years, American Society of Anesthesiologists (ASA) physical status I or II requiring sedation for elective outpatient brain Magnetic Resonance Imaging (MRI) examination were included in the study. ⋯ This study demonstrates that both infusion techniques preserve haemodynamic stability and are associated with minimal complications. The use of an infusion pump, which consistently maintains accurate dosing, reduces the amount of propofol judged adequate by the anaesthesiologist to achieve sedation in MRI. This leads to a more consistently faster emergence and early discharge after sedation in children undergoing MRI studies.
-
Comparative Study
Contactless respiratory and heart rate monitoring: validation of an innovative tool.
To assess the accuracy of the EverOn™ piezoelectric sensor based contactless heart rate and respiration rate monitoring system. ⋯ The system described was found to be accurate in accordance with regulatory and industry criteria.