Journal of medical engineering & technology
-
An abnormal respiratory rate is often the earliest sign of critical illness. A reliable estimate of respiratory rate is vital in the application of remote telemonitoring systems, which may facilitate early supported discharge from hospital or prompt recognition of physiological deterioration in high-risk patient groups. ⋯ Analysis of the photoplethysmogram (PPG) waveform offers an alternative means of non-invasive respiratory rate monitoring, but further development is required to enable reliable estimates. This review conceptualizes the challenge by discussing the effect of respiration on the PPG waveform and the key physiological mechanisms that underpin the derivation of respiratory rate from the PPG.
-
We have designed, built and tested a novel spinal cord surrogate that mimics the low-amplitude cardiac-driven pulsations of the human spinal cord, for use in developing intradural implants to be used in a novel form of neuromodulation for the treatment of intractable pain and motor system dysfunction. The silicone surrogate has an oval cross section, 10 mm major axis × 6 mm minor axis, and incorporates a 3 mm diameter × 3 cm long angioplasty balloon that serves as the pulsation actuator. When pneumatically driven at 1 Hz and 1.5 atmospheres (≈ 1140 mm Hg), the surrogate's diametric pulsation is ≈ 100 μm, which corresponds well to in vivo observations. The applications for this surrogate are presented and discussed.