Pacing and clinical electrophysiology : PACE
-
Pacing Clin Electrophysiol · Apr 2012
Randomized Controlled TrialNear elimination of ventricular pacing in SafeR mode compared to DDD modes: a randomized study of 422 patients.
SafeR performance versus DDD/automatic mode conversion (DDD/AMC) and DDD with a 250-ms atrioventricular (AV) delay (DDD/LD) modes was assessed toward ventricular pacing (Vp) reduction. ⋯ In this selected patient population, SafeR markedly suppressed unnecessary Vp compared with DDD modes.