The Journal of hospital infection
-
Randomized Controlled Trial
A prospective clinical trial to evaluate the microbial barrier of a needleless connector.
Needleless connectors are being increasingly used for direct access to intravascular catheters. However, the potential for microbial contamination of these devices and subsequent infection risk is still widely debated. In this study the microbial contamination rate associated with three-way stopcock luers with standard caps attached was compared to those with Y-type extension set luers with Clearlink needleless connectors attached. ⋯ The internal surfaces of 20 of 200 (10%) three-way stopcock luers with standard caps were contaminated with micro-organisms whereas only one of 193 (0.5%) luers with Clearlink attached was contaminated (P<0.0001). These results demonstrate that the use of the Clearlink device with a dedicated disinfection regimen reduces the internal microbial contamination rate of CVC luers compared with standard caps. The use of such needle-free devices may therefore reduce the intraluminal risk of catheter-related bloodstream infection and thereby supplement current preventive guidelines.