Magnetic resonance imaging
-
The objective of this study was to evaluate the image quality of a respiratory-triggered T2-weighted (T2w) turbo spin-echo (TSE) sequence for magnetic resonance cholangiopancreatography (MRCP) using a new method for respiratory triggering by tracking the motion of the right diaphragm [prospective acquisition correction (PACE) technique]. ⋯ Respiratory-triggered MRCP using a T2w TSE sequence with PACE significantly improves image quality and may be included into the routine MRCP sequence protocol.
-
A controversy has existed over the requirement to cardiac gate diffusion-weighted MRI acquisitions of the brain. Conventional wisdom suggests gating to be a necessary requirement to allow acquisition of accurate data, but recent applications find gating not necessary. The signal-to-noise and acquisition duration of these two approaches can be quite different; thus, this difference in methodology is important. ⋯ Zero padding of the acquisition matrix provides an alternative reconstruction method that is not found to accentuate the artifacts that are due to pulsatile motion in the diffusion-weighted acquisition and thus do not require a gated acquisition. The relative merits of each reconstruction approach are discussed, including estimates of the relative signal-to-noise ratio and resolution benefits. It is concluded that both gated methods and non-gated methods can each provide high quality results with appropriate reconstruction methods.