IEEE transactions on medical imaging
-
IEEE Trans Med Imaging · Nov 2013
Evaluation and real-time monitoring of data quality in electrical impedance tomography.
Electrical impedance tomography (EIT) is a noninvasive method to image conductivity distributions within a body. One promising application of EIT is to monitor ventilation in patients as a real-time bedside tool. Thus, it is essential that an EIT system reliably provide meaningful information, or alert clinicians when this is impossible. ⋯ Additionally, we show that q may be used to compare the performance of EIT systems using phantom measurements. Results suggest that the calculated metric reflects well the quality of reconstructed EIT images for both phantom and clinical data. The proposed measure can thus be used for real-time assessment of EIT data quality and, hence, to indicate the reliability of any derived physiological information.