The Venous Window Needle Guide, a hemodialysis cannulation device for salvage of uncannulatable arteriovenous fistulas.
Arteriovenous fistulas (AVFs) are recommended for hemodialysis access when possible. A noncannulatable but otherwise well functioning AVF leads to prolonged catheter dependency and frustration for the patient and the renal health care provider team. Difficult cannulation patients include obese individuals in whom cannulation sites are too deep, and others with vein segments that are short, tortuous, or otherwise difficult to palpate. The Venous Window Needle Guide for Salvage of AV Fistulae (SAVE) trial was designed to evaluate the efficacy and safety of the Venous Window Needle Guide (VWING; Vital Access Corp, Salt Lake City, Utah) device for salvage of such noncannulatable AVFs that are otherwise functional. ⋯ The VWING was safe and effective in facilitating AVF cannulation for patients with an otherwise mature but noncannulatable fistula. Successful AVF access was achieved using the VWING in 49 (96%) of the 51 patients evaluated. The AVF survival rate at 6 months was 100%.