Canadian journal of anaesthesia = Journal canadien d'anesthésie
-
Multicenter Study
Early clinical experience with a new videolaryngoscope (GlideScope) in 728 patients.
To evaluate a new videolaryngoscope and assess its ability to provide laryngeal exposure and facilitate intubation. ⋯ GS laryngoscopy consistently yielded a comparable or superior glottic view compared with DL despite the limited or lack of prior experience with the device. Successful intubation was generally achieved even when DL was predicted to be moderately or considerably difficult. GS was abandoned in 3.7% of patients. This may reflect the lack of a formal protocol defining failure, limited prior experience or difficulty manipulating the endotracheal tube while viewing a monitor.