The Journal of arthroplasty
-
Corrosion at the modular neck-stem taper junction has become an increasingly important topic as several reports have identified this couple as a possible source for early failure with findings similar to failed metal-on-metal hip arthroplasties. Recently, two different modular stem systems from a single manufacturer were voluntarily recalled due to concerns of failure of the modular taper junction. ⋯ We further reviewed the literature to evaluate whether this is a manufacturer-specific defect or indicative of a broader trend. Recent studies appear to implicate the basic design of the neck-stem taper junction, rather than a single manufacturer, which is at high risk for fretting and corrosion.