Validation of the noninvasive assessment of central blood pressure by the SphygmoCor and Omron devices against the invasive catheter measurement.
To investigate the accuracy of the SphygmoCor and Omron HEM9000-AI devices in the estimation of central blood pressure (BP) in comparison with the simultaneous invasive catheter measurement. ⋯ Both devices underestimated central systolic BP, with a larger deviation by SphygmoCor. Nonetheless, these noninvasive estimations of central BP closely correlate with the invasive measurements, and can still be properly used, on the condition that device specific diagnostic thresholds become available.