Randomized Controlled Trial Comparative StudyComparison of the pharyngeal pressure provided by two heated, humidified high-flow nasal cannulae devices in premature infants.
This study aims to determine if there is a difference in the pharyngeal pressure, measured as a surrogate for continuous positive distending airway pressure, delivered to premature infants between two commonly used heated, humidified high-flow nasal cannulae (HHHFNC) devices: Fisher & Paykel Healthcare HHHFNC and Vapotherm 2000i. ⋯ Both HHHFNC delivered similar pharyngeal pressures at flow rates of 2-6 L/min. The pressure limiter valve of the Fisher & Paykel device attenuated the pharyngeal pressures at flows of 7 and 8 L/min. Vapotherm trended towards higher delivered pharyngeal pressure at flow rates 7 and 8 L/min, but the clinical significance of the difference remains unclear.