Comparison of piezoresistive sensor to PicoPress® in in-vitro interface pressure measurement.
Objective Interface pressure, the sine qua non for compression therapy, is rarely measured in clinical practice and scientific research. The goal of this study aimed to compare and examine the accuracy between a commercially available piezoresistive sensor and PicoPress® (Microlab, Padua, Italy) using the cylinder cuff model to measure in-vitro interface pressure. Method Ten piezoresistive sensors were calibrated using the National Institute of Standard and Technology certified manometer, and compared to PicoPress® using cylinder cuff model from 20 to 120 mmHg. ⋯ Using mean pressure value, there was no significant difference between the front surface, back surface of the piezoresistive sensors, and PicoPress® (P > 0.05). Standard deviation was larger for the piezoresistive sensors than PicoPress® at any given pressure and this difference was more pronounced in the higher pressure range. Conclusion Piezoresistive sensor may represent a viable alternative to PicoPress® in interface pressure measurement.