International journal of hematology
-
Disseminated intravascular coagulation (DIC) diagnosis is hampered by the limited availability of reliable clinical or laboratory tests. Currently available tests are time consuming and expensive. We investigated whether coagulation and platelet function analyses using the Sonoclot system were suitable for overt DIC diagnosis in critically ill adults. ⋯ Significant correlations were observed between activated clotting time (ACT) and ISTH score (r = 0.7; P < 0.001), clot rate (CR) and ISTH score (r = 0.5; P < 0.001), platelet function (PF) and ISTH score (r = -0.6; P < 0.001), and PF and platelet count (r = 0.5; P < 0.001). An ACT cut-off value of 213.5 s alone or combined with CR presented good sensitivity (76.7 and 86.8 %, respectively) and specificity (96.2 and 93.3 %, respectively). Sonoclot analysis can be performed using a point-of-care device that effectively discriminates low and high ISTH scores, and that effectively predicts coagulation dysfunction in patients with overt DIC.