ASAIO journal : a peer-reviewed journal of the American Society for Artificial Internal Organs
-
Cardiopulmonary bypass (CPB) devices and techniques have continuously evolved. We have conducted surveys that chronicle the changes in CPB devices and techniques used at North American pediatric cardiac surgery centers since 1989. The aim of this article is to describe trends in cardiopulmonary bypass device use during cardiac surgery and changes in the devices used for extracorporeal life support (ECLS) following cardiac surgery for pediatric patients. ⋯ ECLS systems comprised of hollow fiber oxygenators and centrifugal pumps for are gradually replacing the classical ECLS circuit, servo regulated roller pumps and silicone rubber membranes. Nearly 40% of centers use these alternate components in their ECLS systems. Costs, utility, safety and measurable benefit to the patient should guide decisions related to device selection.
-
Blood gas management during hypothermic cardiopulmonary bypass may be corrected by pH stat or alpha stat strategy. The pH stat philosophy is to maintain the blood pH constant at any temperature. Carbon dioxide must be introduced to the oxygenator in order to maintain the pH and pCO2 during hypothermic cardiopulmonary bypass. ⋯ Results from several studies favor the pH stat strategy during neonatal cardiopulmonary bypass. This strategy increases tissue oxygenation and cerebral blood flow while cooling. Data also suggest that pH stat management results in better outcomes with shorter ventilation times and intensive care unit stays after pediatric cardiac surgery.