Effects of lumbar artificial disc design on intervertebral mobility: in vivo comparison between mobile-core and fixed-core.
Although in theory, the differences in design between fixed-core and mobile-core prostheses should influence motion restoration, in vivo kinematic differences linked with prosthesis design remained unclear. The aim of this study was to investigate the rationale that the mobile-core design seems more likely to restore physiological motion since the translation of the core could help to mimic the kinematic effects of the natural nucleus. In vivo intervertebral motion characteristics of levels implanted with the mobile-core prosthesis were compared with untreated levels of the same population, levels treated by a fixed-core prosthesis, and normal levels (data from literature). ⋯ Furthermore, some physiologic mechanical behaviors seemed to be maintained: the VT was higher at implanted the L4L5 level than at the implanted L5S1 level, and the CT appeared lower at the L4L5 level than at the L5S1 level. ROM and motion distribution were not different between the mobile-core prosthesis and the fixed-core prosthesis implanted levels. This study validated in vivo the concept that a mobile-core helps to restore some physiological mechanical characteristics of the VT at the implanted L4L5 level, but also showed that the minimizing effect of core mobility on the VT was not sufficient at the L5S1 level.