Is MAGEC X better than earlier designs of magnetically controlled growing rod: an explant study.
Determine the performance of MAGEC X rods through retrieval analysis and comparison with clinical data. ⋯ Despite the substantial design changes with MAGEC X tissue metallosis was seen in most cases; most explanted MAGEC X rods had lengthened only partially and produced no force output. While the previous issue of locking pin fracture appears to have been mitigated with MAGEC X, there are multiple other and new failure modes, such as endcap separation. Overall MAGEC X appears to give little improvement over earlier iterations of the rod.