Beam hardening filters have long been employed in X-ray Computed Tomography (CT) to preferentially absorb soft and low-energy X-rays having no or little contribution to image formation, thus allowing the reduction of patient dose and beam hardening artefacts. In this work, we studied the influence of additional copper (Cu) and aluminium (Al) flat filters on patient dose and image quality and seek an optimum filter thickness for the GE LightSpeed VCT 64-slice CT scanner using experimental phantom measurements. Different thicknesses of Cu and Al filters (0.5-1.6mm Cu, 0.5-4mm Al) were installed on the scanner's collimator. ⋯ An added Cu filter with approximately 0.5mm thickness accounts for 50% reduction in radiation-absorbed dose as measured by the weighted CTDI. The FOM results indicate that with an additional filter of 0.5mm Cu or minimum 4mm Al, a good compromise between image quality and patient dose is achieved for CT images acquired at tube voltages of 120 and 140kVp. The results seem to indicate that an optimum filter for high kVp acquisitions, routinely used in cardiovascular imaging, should be 0.5mm copper or 4mm aluminium minimum.