Randomized Controlled Trial Comparative Study Clinical TrialVariability in time delay between two models of pulse oximeters for deriving the photoplethysmographic signals.
Pulse oximetry is commonly used as an arterial blood oxygen saturation (SaO2) measure. However, its other serial output, the photoplethysmography (PPG) signal, is not as well studied. Raw PPG signals can be used to estimate cardiovascular measures like pulse transit time (PTT) and possibly heart rate (HR). ⋯ Novametrix differed from ECG by 0.71+/-0.58% (p<0.05) while Masimo differed by 4.51+/-3.66% (p>0.05). Modern oximeters can be attractive for their improved SaO2 measurement. However, using raw PPG signals obtained directly from these oximeters for timing-related measurements warrants further investigations.