Tracheal intubation following training with the GlideScope compared to direct laryngoscopy.
Tracheal intubation using direct laryngoscopy has a high failure rate when performed by untrained medical personnel. This study compares tracheal intubation following direct laryngoscopy by inexperienced medical students when initially trained by using either the GlideScope, a video assisted laryngoscope, or a rigid (Macintosh) laryngoscope. Forty-two medical students with no previous experience in tracheal intubation were randomly divided into two equal groups to receive training with the GlideScope or with direct laryngoscopy. ⋯ The rates of successful intubation were significantly higher in the Glidescope group after the first (48%), second (62%), and third (81%) intubations compared with the Macintosh group (14%, 14% and 33%; p = 0.043, 0.004 and 0.004, respectively). The mean (SD) times for the first, second, and third successful tracheal intubations were significantly shorter in the Glidescope group (59.3 (4.4) s, 56.6 (7.1) s and 50.1 (4.0) s) than the Macintosh group (70.7 (7.5) s, 73.7 (7.3) s and 67.6 (2.0) s; p = 0.006, 0.003 and 0.0001, respectively). Training with a video-assisted device such as the GlideScope improves the success rate and time for tracheal intubation in patients with normal airways when this is performed by inexperienced individuals following a short training programme.
-
Unanticipated difficulties during tracheal intubation and failure to intubate are among the leading causes of anaesthesia-related morbidity and mortality. Using the technique of video laryngoscopy, the alignment of the oral and pharyngeal axes to facilitate tracheal intubation is unnecessary. In this study we evaluated the McGrath Series 5 videolaryngoscope for tracheal intubation in 61 patients who exhibited Cormack and Lehane grade 3 or 4 laryngoscopies with a Macintosh laryngoscope. ⋯ Laryngoscopy was improved by one grade in 10%, by two grades in 80% and by three grades in 10% of cases (p < 0.0001). The success rate for intubation was 95% with the McGrath. These results suggest that the McGrath videolaryngoscope can be used with a high success rate to facilitate tracheal intubation in difficult intubation situations.
-
Invasive ventilation poses a risk of respiratory infection that can be drug-resistant. One means of reducing transmission of infection is the use of a breathing system filter. Filters are intended to be used with dry gas. ⋯ All the filters tested allowed free passage of both Candida albicans and coagulase-negative staphylococci. The median (IQR [range]) pressure required for fluid to flow across the filter varied greatly between different filter types (20 (0-48 [0-138]) cmH(2)O). We conclude that even large microorganisms pass across moist breathing system filters in conditions that are found in clinical practice.