Comparison of portable blood-warming devices under simulated pre-hospital conditions: a randomised in-vitro blood circuit study.
Pre-hospital transfusion of blood products is a vital component of many advanced pre-hospital systems. Portable fluid warmers may be utilised to help prevent hypothermia, but the limits defined by manufacturers often do not reflect their clinical use. The primary aim of this randomised in-vitro study was to assess the warming performance of four portable blood warming devices (Thermal Angel, Hypotherm X LG, °M Warmer, Buddy Lite) against control at different clinically-relevant flow rates. ⋯ There was no association between warming and haemolysis with any device (p = 0.949) or flow rate (p = 0.169). Practical issues, which may be relevant to clinical use, also emerged during testing. Our results suggest that there were significant differences in the performance of portable blood warming devices used at flow rates encountered in clinical practice.
-
The incidence of an anaesthetic drug error can be directly observed in large trials. In an alternative approach, we developed a probabilistic mathematical model in which the anaesthetist is modelled as a 'fallible entity' who makes repeated drug administration choices during an operation. This fallibility was factored in the model as an initial 'intrinsic error rate'. ⋯ The practice implications of our modelling include: exercising caution or avoiding starting work if under par; added vigilance in unfamiliar environments; keeping anaesthetic recipes simple; and recognising that operation durations > 5-6 h constitute a time of exaggerated risk. These implications are testable predictions in observational trials. If validated, our model could serve as a potential research tool to investigate the impact of safety interventions on the rate of intrinsic error using simulation.
-
Phase lag entropy, an electro-encephalography-based hypnotic depth indicator, calculates diversity in temporal patterns of phase relationship. We compared the performance of phase lag entropy with the bispectral index™ in 30 patients scheduled for elective surgery. We initiated a target-controlled infusion of propofol using the Schnider model, and assessed sedation levels using the Modified Observer's Assessment of Alertness/Sedation scale every 30 s with each stepwise increase in the effect-site propofol concentration. ⋯ Baseline variability in phase lag entropy and bispectral index was 3.3 and 5.7, respectively. The predicted propofol concentrations, using the Schnider pharmacokinetic model, producing a 50% probability of moderate and deep sedation were 1.96 and 3.01 μg.ml-1 , respectively. Phase lag entropy was found to be useful as a hypnotic depth indicator in patients receiving propofol sedation.
-
Excessive noise in hospitals adversely affects patients' sleep and recovery, causes stress and fatigue in staff and hampers communication. The World Health Organization suggests sound levels should be limited to 35 decibels. This is probably unachievable in intensive care units, but some reduction from current levels should be possible. ⋯ Some of this was likely to be appropriate communication between the patient, their caring staff and visitors. However, a significant proportion of loud sounds may originate from equipment alarms which are sited at the bedside. A redesign of the intensive care unit environment to move alarm sounds away from the bed-side might significantly reduce the environmental noise burden to patients.
-
Editorial Comment
Dexamethasone in regional anaesthesia: travelling up a blind alley?